 LINEEYE Software for Transferring Firmware   Version 1.04
                    << le8firm.exe >>
                                           (C)2011, 2008 by LINEEYE CO.,LTD.
****************************************************************************

============================================================================
1. Outline
    Thank you for purchasing the LINEEYE portable communication analyzer.
    This software is for updating the firmware of your analyzer from the PC.

============================================================================
2. Appropriate models
      LE-8200/ LE-3500/ LE-2500/ LE-1500/ LE-270A
    * LE-200PS/ LE-150PS Utility software contains the function of
      transferring firmware.   

============================================================================
3. Preparation

3-1. If using the AUX port for connection (only for LE-8200 / LE-3500 /
     LE-2500 / LE-1500)
    (1) Connect the AUX port of your analyzer and the COM port of the PC.

    (2) Set the communication conditions of the AUX port of your analyzer.
        Speed:115200/230400, Data bit:8, Parity:None, X-control:OFF
        (Set the same speed as PC setting.)

    (3) Execute the "Le8firm.exe" and set "Serial port" and "baudrate".
        "Serial port" : Select the COM port (COM1-COM99) of the PC which
                        connects to your analyzer.
        "baudrate"    : Set the same communication speed being set in your
                        analyzer.

3-2. If using the USB port for the connection
    (1) Install the USB driver in the PC
        (Refer to the instruction manual of analyzer for details of
        installing the USB driver.)

    (2) Le8firm 
        Execute "Le8firm.exe". Select "USB" for connection and input the
        serial number. 


============================================================================
4. Transfer the firmware

   Install the latest firmware from our webpage.

    (1) Analyzer
      Turn off the power of your analyzer.
        * LE-8200
          Turn on the power of your analyzer while pressing
          [Shift] and [File]
        * LE-3500, LE-2500, LE-1500
          Turn on the power of your analyzer while pressing
          [SHIFT] and [LOAD/SAVE].
        * LE-270A
          No need to operate.

    (2) Le8firm 
      1) Execute "Le8firm.exe". Select the type of connection and other
         settings, such as COM port number and serial number.
      2) ClickmNext].
      3) Click [Select] and select the firmware file (...FW2)
      4) Click [Next].
      5) Click [Start]and start transferring the firmware.
      6) "Complete" message will be appeared.
      7) Click [Close].

****************************************************************************
*Connect the AC adapter to the analyzer before starting the operation      *
*Never turn off the power of your analyzer while transferfing the firmware.*
*Otherwise, transferfing program may not complete the operation correctly  *
*and you may need to send back the analyzer to LINEEYE.                    *
****************************************************************************

    (3) Turn off the power of your analyzer. Then turn on the power of your
        analyzer again and it will run by the latest firmware.

============================================================================
5. Environment

    Windows 2000/ XP/ Vista/ 7/ 8

    *  It will display in Japanese if you are using Japanese Windows.
       Anything else will display in English.

============================================================================
6. Version History
    
    Version 0.21  First version  (May, 2008)
    Version 1.00  Correct the minor bug about the Japanese and English
                  display.
    Version 1.02  Support LE-2500 and LE-3500. (October, 2008)
    Version 1.04  Support LE-270A. (November, 2011)

============================================================================

LINEEYE CO., LTD.        URL      http://www.lineeye.com
                         E-mail   trade@lineeye.co.jp  
                                                                M-538UT1E/LE
============================================================================